AUTO GBM-L2 Dual-Pressure Methods Vacuum Decay Leak Detection Tester adopts the test principle of vacuum decay method/pressure decay to detect micro-leakage of finished packaging, It can be applied to packaging systems composed of vials, ampoules, pre-filled syringes, eye drop bottles, HDPE bottles, infusion bottles/bags and other pharmaceutical packaging, food packaging, and chemical industry packaging.
| Absolute Pressure Testing Range | (0-300) kPa |
| Differential Pressure Testing Range | (-2-2) kPa |
| Detection Sensitivity | 1-3 um |
| Balancing / Testing Time | 1-3600 s |
| Vacuum Purge Time | 1-3600 s |
| Set Flow Rate | 0-3 mL/min |
| Testing System | Dual Sensor Technology / Dual Cycle Testing |
| Testing Chamber | Customized according to the sample |
| Applicable Products | Vials, Ampoules, Prefilled Syringes, Infusion Bottles/Bags, etc. |
| Detection Principle | Vacuum Decay Method / Pressure Decay Method |
| Dimensions of the Main Unit | 495 mm*455 mm*280 mm |
| Ambient Temperature | 20°C-25°C |
| Relative Humidity | 40%-60% |
| Working Power Supply | AC 220V, 50HZ |
No brochures available.